电感耦合等离子体质谱法测定医用镍钛形状记忆合金中钴、铜、铬、铁、铌
王长华12,王伟华3,墨淑敏1,胡芳菲1,李爱嫦1
1.国标(北京)检验认证有限公司,北京 101400; 2.国合通用测试评价认证股份公司,北京 101400; 3.北京环标科创环境科技发展有限责任公司,北京 100029 
Determination of cobalt, copper, chromium, iron and niobium in medical nickel titanium shape-memory alloy by inductively coupled plasma-mass spectrometry
WANG Changhua12, WANG Weihua3, MO Shumin1, HU Fangfei1, LI Aichang1
1. Guobiao (Beijing) Testing & Certification Co., Ltd., Beijing 101400, China; 2. China United Test & Certification CO., Ltd., Beijing 101400, China; 3. Beijing Huan biao Ke chuang Environmental Technology Development Co., Ltd., Beijing 100029, China 
摘要:建立电感耦合等离子体质谱法测定医用镍钛形状记忆合金中钴、铜、铬、铁和铌5种杂质元素含量的方法。采用硝酸-氢氟酸混合酸溶解样品,选择H2动态反应池模式消除干扰,以Cs为内标进行校正。钴、铜、铬、铁、铌的质量浓度在0~100 ng/mL范围内与离子响应信号具有良好的线性关系,相关系数均大于0.999,检出限为0.036~0.13 μg/g,测定下限为0.12~0.44 μg/g。用所建方法对实际样品进行测定,测定结果的相对标准偏差为1.2%~2.7%(n=7),样品加标回收率为97.0%~102.2%。该方法操作简单,干扰少,满足医用镍钛合金类产品中杂质元素含量的测定要求。  
Abstract:A method for the determination of cobalt, copper, chromium, iron and niobium in medical nickel titanium shape-memory alloy by inductively coupled plasma-mass spectrometry(ICP-MS) was established. Nitric acid hydrofluoric acid mixed acid was used to dissolve the sample, the interference was eliminated by H2 dynamic reaction tank mode, and the calibration was carried out with Cs as internal standard. The mass concentration of Co, Cu, Cr, Fe, Nb had a good linear relationship with the ion response signal in the range of 0.5-100 ng/mL, the correlation coefficients were all more than 0.999, the detection limits were 0.036-0.13 μg/g, and the lower limits of determination were 0.12-0.44 μg/g. The method was applied to the determination of real samples, the relative standard deviations of the determination results were 1.2%-2.7%(n=7), and the recoveries were 97.0%-102.2%. The method is simple in operation, less interference, which can meet the requirements of determination of impurity elements in medical nickel titanium alloy products.  
关键词:电感耦合等离子体质谱法;医用镍钛形状记忆合金;杂质元素 
Keywords:inductively coupled plasma-mass spectrometry; medical nickel-titanium shape-memory alloy; impurity elements 
基金:工信部国家新材料测试评价平台主中心项目(52801) 
本文引用格式:
王长华,王伟华,墨淑敏,等.电感耦合等离子体质谱法测定医用镍钛形状记忆合金中钴、铜、铬、铁、铌[J].化学分析计量,2020,29(6):47-50.
WANG C H,WANG W H,MO S M,et al. Determination of cobalt,copper,chromium,iron and niobium in medical nickel titanium shape-memory alloy by inductively coupled plasma-mass spectrometry[J]. Chemical analysis and meterage,2020,29(6):47-50. 

全文:
相似文章:
  电感耦合等离子体质谱法测定阿莫西林原料药中的7种杂质元素
2021年第06期
  微波消解-ICP-MS法测定车用金属催化剂中铂、钯、铑
2021年第06期
  电感耦合等离子体质谱法测定土壤中有效钼
2021年第05期
  高效液相色谱-电感耦合等离子体串联质谱法测定水样中6种硒形态
2021年第05期
  微波消解-电感耦合等离子体质谱法测定依度沙班原料药中36种金属元素
2021年第04期
  微波消解-电感耦合等离子体质谱法测定铀合金中铌、钼、锆
2021年第02期
  微波消解-电感耦合等离子体质谱法测定酱卤肉中铬、砷、镉、铊、铅
2021年第01期
  电感耦合等离子体质谱法检测农作物土壤中的镉、锡、锰、锑、钒、钛、锶金属污染物残留
2021年第01期
  电感耦合等离子体质谱法测定二氧化铀-氧化铍芯块中28种杂质元素
2020年第04期
  电感耦合等离子体质谱法测定替比培南酯颗粒剂中铑和钯
2020年第04期
《化学分析计量》杂志社 版权所有,未经授权禁止复制或者建立镜像
地址:山东省济南市天桥区田庄东路3号 邮编:250031
电话:0531-85878132 85878148 85878224 85878278 传真:0531-85947355
鲁ICP备05018524号-6