《化学分析计量》 2020年第02期
DOI:10.3969/j.issn.1008-6145.2020.02.028
电感耦合等离子体质谱法测定食用植物油中铅、总砷含量不确定度评定
李赵平
开封市食品药品检验所,河南开封 475000 
Uncertainty evaluation in determination of lead and arsenic in edible vegetable oil by inductively coupled plasma mass spectrometry
李赵平
Kaifeng Center for Food and Drug Control, Kaifeng 475000, China 
摘要:采用电感耦合等离子体质谱法(ICP-MS) 测定食用植物油中铅(Pb)、总砷(As) 含量,对测定结果的不确定度进行评定。利用ICP-MS,采用标准加入法测定食用植物油中Pb,As 含量,分析测定过程中不确定度来源,包括样品称量、定容、前处理、溶液中Pb 和As 浓度、重复性测量等引入的不确定度,计算合成不确定度。结果表明,Pb 含量为(0.022 8±0.013) mg/kg,k=2 ;As 含量为(0.007 85±0.004 5) mg/kg,k=2。该方法的不确定度主要来源于样品溶液中的Pb,As 浓度,评定得到的不确定度可为正确评价测定结果提供科学依据。 
Abstract:The content of lead (Pb) and arsenic (As) in edible vegetable oil was determined by inductively coupled plasma mass spectrometry (ICP-MS). The uncertainty of the determination results was evaluated. The content of Pb and As in edible vegetable oil was determined by ICP-MS using standard addition method, The sources of uncertainty in the determination process were analyzed, including the uncertainty introduced by sample weighing, constant volume, pretreatment, concentration of Pb and As in solution, repeatability measurement, etc. the synthetic uncertainty was calculated. The results showed that the Pb content was (0.022 8±0.013) mg/kg, k=2, and the As content was (0.007 85±0.004 5) mg/kg, k=2. The uncertainty of the method is mainly the concentration of Pb and As in the sample solution. The uncertainty obtained by the evaluation can provide a scientific basis for the correct evaluation of the measurement results. 
关键词:不确定度;电感耦合等离子体质谱法;食用植物油;铅;总砷 
Keywords:uncertainty; ICP-MS; edible vegetable oil; lead; total arsenic 
本文引用格式:
李赵平. 电感耦合等离子体质谱法测定食用植物油中铅、总砷含量不确定度评定[J]. 化学分析计量,2020,29(2):118-122.
LI Z P. Uncertainty evaluation in determination of lead and arsenic in edible vegetable oil by inductively coupled plasma mass spectrometry[J] . Chemical analysis and meterage,2020,29(2):118-122. 

全文:
相似文章:
  Top-down控制图法评定石墨炉原子吸收光谱法测定血中铅含量的不确定度
2024年第02期
  气相色谱法测定酱油中对羟基苯甲酸甲酯的不确定度评定
2024年第02期
  电感耦合等离子体质谱法测定复杂高温合金中微量硅
2024年第02期
  电感耦合等离子体质谱法测定甜叶菊中25种元素
2024年第02期
  Top-down法评定EDTA络合滴定法测定海洋沉积物中碳酸钙含量的不确定度
2024年第01期
  超高效液相色谱-串联质谱法检测蘑菇中6种蘑菇毒素的不确定度评定
2024年第01期
  电感耦合等离子体质谱法测定纹身贴中17种有害元素及其迁移量
2024年第01期
  基于WPSVBA编程的Top-Down法评定化学分析测量不确定度
2023年第11期
  电感耦合等离子体原子发射光谱法测定药用缓释材料中锡含量的不确定度评定
2023年第11期
  一种溶解氧测定仪自动检定装置的研制
2023年第11期
《化学分析计量》杂志社 版权所有,未经授权禁止复制或者建立镜像
地址:山东省济南市天桥区田庄东路3号 邮编:250031
电话:0531-85878132 85878224 85878278
鲁ICP备05018524号-6