收藏我们   ENGLISH  
论文查询
范围:  
提 交
《化学分析计量》 2017年第05期
DOI:10.3969/j.issn.1008-6145.2017.05.015
基于钼酸盐光度法测定总磷的浊度去除方法
孙娟,徐荣,卢秋云,乔丹丹
南京市环境监测中心站,南京 210013 
Turbidity Removal Method of Total Phosphorus Determination Based on Molybdate Photometry
孙娟,徐荣,卢秋云,乔丹丹
Nanjing Environmental Monitoring Center Station, Nanjing 210013, China 
摘要:采取新的浊度去除方法,改进了测定总磷的钼酸盐光度法。测定过程中,消解液的浊度会对测定结果产生干扰。通过比较机械过滤法和试剂补偿法的浊度校正效果,建立了中速定性滤纸单次机械过滤的浊度去除方法,确定了最佳去除条件:控制待滤液温度不高于30℃,清洗纯水的体积不少于20 mL。选择消解液具有浊度干扰项的6种代表性水样,分别进行平行测定和加标回收试验,方法检出限为0.008 mg/L,测定结果的相对标准偏差为2.8%~7.9%(n=6),加标回收率为92.0%~106%。该方法灵敏度、准确度高,精密度良好,与现行国标法相比,试剂消耗量少,分析效率高,能够满足多种水环境样品中总磷的测定要求。 
Abstract:A turbidity removal method for total phosphorus determination was established based on molybdate photometry. The turbidity in the dissolving fluid interfered with the result of the detection. Turbidity correction effect of mechanical filtering method was compared with that of reagent compensation method, and then, a single mechanical filtration turbidity removal of medium-speed qualitative filter method had been established. At the same time, the best removal condition was identified, which included that the filtrate temperature was not more than 30℃and cleaning water volume was not less than 20 mL. Six kinds of typical selected samples which their dissolving liquid had turbidity interference term were tested for parallel and additive recovery measurements. The detection limit of the method was 0.008 mg/L, the relative standard deviation was 2.8%-7.9%(n=6), and the recoveries ranged from 92.0% to 106%. The method has high sensitivity, accuracy and precision. Compared with the current national standard method, it has advantages including lower reagent consumption and higher analytical efficiency, which can meet the detection requirement of total phosphorus in a variety of samples in the water environment. 
关键词:钼酸盐光度法;总磷;浊度;去除方法 
Keywords:molybdate photometry; total phosphorus; turbidity; removal method 
本文引用格式:
孙娟,徐荣,卢秋云,等.基于钼酸盐光度法测定总磷的浊度去除方法[J].化学分析计量,2017,26(5):59-62. 

全文:
相似文章:
  标准曲线质量控制图在水质总磷检测中应用
2022年第10期
  总磷测定仪校准方法
2021年第01期
  比色法总磷自动监测仪的校准
2020年第06期
  国家监测网方法测定地表水中总磷与国标法差异及改进
2020年第01期
  连续流动法同时测定污水中氨氮和总磷
2020年第01期
  区带流动分析技术测定水中的总磷
2019年第05期
  钼酸铵分光光度法测定地表水总磷含量的样品采集与预处理
2019年第01期
  基于壳聚糖的浊度法测定超滤膜截留率
2017年第06期
  水质浊度标准物质的研制
2016年第04期
  紫外消解流动注射光度法测定海水养殖废水中总氮、总磷
2015年第03期
《化学分析计量》杂志社 版权所有,未经授权禁止复制或者建立镜像
地址:山东省济南市天桥区田庄东路3号 邮编:250031
电话:0531-85878132 85878224 85878278
鲁ICP备05018524号-6