收藏我们   ENGLISH  
论文查询
范围:  
提 交
《化学分析计量》 2016年第06期
DOI:10.3969/j.issn.1008-6145.2016.06.008
双柱串联-便携式气相色谱法测定室内空气中苯系物
潘扬昌,莫达松,黄璐璐,梁勇,谢安杰,覃仁港
广西壮族自治区分析测试研究中心,南宁 530022 
Determination of BTEX in Room Air by Tandem Double Column-portable Gas Chromatography
潘扬昌,莫达松,黄璐璐,梁勇,谢安杰,覃仁港
Guangxi Zhuang Autonomous Region Center for Analysis and Test Research, Nanning 530022, China 
摘要:建立双柱串联-便携式气相色谱测法定室内空气中苯系物的方法。利用仪器内部自带的Tenax吸附剂在0.5 L/min流量下采集空气样品100 s,然后仪器迅速升温至400℃左右,对吸附剂进行热解吸进样,以微亚离子检测器(MAID)对室内空气中7种苯系物进行测定。结果表明,色谱条件为MXT-200/MXT-WAX金属毛细管柱串联色谱柱,柱温90℃,柱压103425 Pa时,分离效果最好,且运行时间短。各组分在19.26~306.7 μg/m3范围内与色谱峰面积的平方根线性关系良好,相关系数r大于0.995。方法检出限为6.40~15.9 μg/m3,空白加标回收率为92%~114%,测定结果的相对标准偏差为6.1%~13.7%(n=6)。采用双柱串联便携式气相色谱法测定室内空气中苯系物具有快速、准确、方便等优点。 
Abstract:Determination of benzene series(BTEX) in indoor air by tandem double column-portable gas chromatography was developed. The indoor air samples were collected 100 s by Tenax adsorbent for instrument in 0.5 L/min flow, and the instrument was quickly heated to 400℃ for thermal desorption injection, and the seven kinds of BTEX were detected by micro argon ionization detector(MAID). The chromatographic conditions were as follows: MXT-200 and MXT-WAX metal capillary chromatographic column in series was used, the column temperature was 90℃, column pressure was 103425 Pa. Under this condition, the separation effect was best, and the operation time was short. The mass concentration of each compounds and the square root of peak area had good linear relationship in the range of 19.26-306.7 g/m3, and the correlation coefficients were more than 0.995. The detection limit was 6.40-15.9 μg/m3, the recovery rate of the blank addition of each component was 92%-114%, and the relative standard deviation of determination results was 6.1%-13.7%(n=6). The method is rapid, accurate and convenient. 
关键词:苯系物;室内空气;便携式气相色谱法;双柱串联;微亚离子检测器 
Keywords:BTEX; indoor air; portable gas chromatography; double column series; micro argon ionization detector 
基金:广西壮族自治区直属公益性科研院所基本科研业务费专项 
本文引用格式:
潘扬昌,莫达松,黄璐璐,等.双柱串联-便携式气相色谱法测定室内空气中苯系物[J].化学分析计量,2016,25(6):34-37 .
Pan Yangchang,Mo Dasong,Huang Lulu,et al.Determination of BTEX in Room Air by Tandem Double Column-portable Gas Chromatography[J]. Chemical analysis and meterage,2016,25(6):34-37. 

全文:
相似文章:
  顶空-气相色谱法测定环境水体中8 种苯系物
2021年第10期
  室内空气中甲醛检测方法研究进展
2021年第08期
  热分离进样-低热容气相色谱-质谱法测定空气中苯系物
2021年第07期
  活性炭吸附/二硫化碳解析-气相色谱法测定环境空气中11种苯系物
2021年第07期
  上海市室内空气质量检测能力验证结果分析
2021年第04期
  热脱附-冷阱捕集-气相色谱-质谱法测定室内空气中16种挥发性有机物
2021年第04期
  顶空气相色谱法测定驱蚊扣中5种苯系物
2020年第04期
  气相色谱法测定密闭空间空气中7种苯系物
2020年第03期
  顶空-气相色谱法测定水中7种苯系物
2019年第06期
  多束毛细管柱-离子迁移率谱检测环境大气中的苯系物
2018年第03期
《化学分析计量》杂志社 版权所有,未经授权禁止复制或者建立镜像
地址:山东省济南市天桥区田庄东路3号 邮编:250031
电话:0531-85878132 85878224 85878278
鲁ICP备05018524号-6